Hylke H A C M van der Heijden1, Gerben J Truin2, Joyce Verhaeg1
1Department of Intensive Care Medicine, Radboud University Medical Centre, Nijmegen, The Netherlands.
Introduction:
Capnography is used to monitor the endtidal carbon dioxide tension (EtCO2) in exhaled gas. Sidestream capnography has great potential to monitor mechanically ventilated pediatric patients, given the continuous sampling from the endotracheal tube into a gas sensor. However, hemodynamic and respiratory impairments may reduce reliability and validity of sidestream capnography to monitor arterial carbon dioxide tension (PaCO2) in critically ill, mechanically ventilated children.
Methods:
In 47 mechanically ventilated pediatric patients (aged 0-14 years, median age 17.2 months), a total of 341 consecutive measurements of PaCO2, EtCO2, respiratory and hemodynamic parameters were performed, and capnogram shape was determined. Validity was assessed with the Bland-Altman limit of agreement (loa), mixed models were used to adjust for variation between patients, and potential confounders were considered with multivariable analyses.
Results:
EtCO2 (mean 4.50 ± 0.96 kPa) underestimated PaCO2 (mean 5.37 ± 0.87) considerably, resulting in a loa of 0.87 kPa [95% confidence interval (95% CI) -1.03;2.77] and 42.2% percentage error. The association improved significantly b = 0.54 [95 %CI = 0.45;0.64, P < 0.001] when corrected for individual differences. The association between EtCO2 and PaCO2 was not influenced by any of the potential confounders.
Conclusions:
Sidestream capnography in mechanically ventilated infants and children seems moderately reliable and valid when corrected for individual differences. Therefore, it could only be used with caution for trend estimation in the individual patient.
More Related Videos
07:51Evaluation of Capnography Sampling Line Compatibility and Accuracy when Used with a Portable Capnography Monitor
Published on: September 29, 2020
08:25Combining Volumetric Capnography And Barometric Plethysmography To Measure The Lung Structure-function Relationship
Published on: January 8, 2019
Related Concept Videos
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Chronic Obstructive Pulmonary Disease-IV: Assessement and Diagnostic Studies
Medical History
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Acute Respiratory Failure-III
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...