Novel Pulse Oximetry Sonifications for Neonatal Oxygen Saturation Monitoring: A Laboratory Study
Kelly Hinckfuss1, Penelope Sanderson2, Robert G Loeb3
1The University of Queensland, St. Lucia, AustraliaUniversity of Arizona, TucsonMater Mothers' Hospital, Brisbane, AustraliaThe University of Queensland, St. Lucia, Australia.
Objective:
We aimed to test whether the use of novel pulse oximetry sounds (sonifications) better informs listeners when a neonate's oxygen saturation (SpO2) deviates from the recommended range.
Background:
Variable-pitch pulse oximeters do not accurately inform clinicians via sound alone when SpO2 is outside the target range of 90% to 95% for neonates on supplemental oxygen. Risk of blindness, organ damage, and death increase if SpO2 remains outside the target range. A more informative sonification may improve clinicians' ability to maintain the target range.
Method:
In two desktop experiments, nonclinicians' ability to detect SpO2 range and direction of change was tested with novel versus conventional sonifications of simulated patient data. In Experiment 1, a "shoulder" sonification used larger pitch differences between adjacent saturation percentages for SpO2 values outside the target range. In Experiment 2, a "beacon" sonification used equal-appearing pitch differences, but when SpO2 was outside the target range, a fixed-pitch reference tone from the center of the target SpO2 range preceded every fourth pulse tone.
Results:
The beacon sonification improved range identification accuracy over the control display (85% vs. 60%; p < .001), but the shoulder sonification did not (55% vs. 52%).
Conclusion:
The beacon provided a distinct auditory alert and reference that significantly improved nonclinical participants' ability to identify SpO2 range.
Application:
Adding a beacon to the variable-pitch pulse oximeter sound may help clinicians identify when, and by how much, a neonate's SpO2 deviates from the target range, particularly during patient transport situations when auditory information becomes essential.
More Related Videos
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
07:51Standardized Hemorrhagic Shock Induction Guided by Cerebral Oximetry and Extended Hemodynamic Monitoring in Pigs
Published on: May 21, 2019
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment: