25-Hydroxyvitamin D assays: Potential interference from other circulating vitamin D metabolites
G D Carter1, J C Jones1, J Shannon1
1Charing Cross Hospital, Fulham Palace Road, London W6 8RF, UK.
The Vitamin D External Quality Assessment Scheme (DEQAS) evaluates 25-hydroxyvitamin D (25-OHD) assays, identifying significant interference from other vitamin D metabolites. This assessment ensures accuracy and specificity in routine laboratory vitamin D testing.
Area of Science:
- Clinical Chemistry
- Endocrinology
- Laboratory Medicine
Background:
- The Vitamin D External Quality Assessment Scheme (DEQAS) is a global program assessing 25-hydroxyvitamin D (25-OHD) measurement accuracy.
- DEQAS expanded in 2015 to include 24,25-dihydroxyvitamin D (24,25(OH)2D) assessment.
- Since 2013, DEQAS 25-OHD assessments are accuracy-based, utilizing the NIST reference measurement procedure (RMP).
Purpose of the Study:
- To evaluate the accuracy and specificity of routine 25-OHD laboratory methods.
- To investigate the interference of other vitamin D metabolites in 25-OHD assays.
- To assess the performance of various analytical methods in resolving different vitamin D metabolites.
Main Methods:
- DEQAS distributes proficiency testing samples to approximately 1100 participants worldwide.
- Samples are spiked with known concentrations of 3-epi-25-OHD3, 24R,25(OH)2D3, and 24S,25(OH)2D3 to assess metabolite interference.
- Analytical methods evaluated include competitive protein binding assays, antibody-based assays, HPLC/UV, and LC-MS/MS.
Main Results:
- The 3-epimer showed 56% cross-reactivity in competitive protein binding assays but was undetectable in antibody-based methods.
- HPLC/UV and LC-MS/MS methods could not always resolve 3-epi-25-OHD3, leading to overestimated total 25-OHD.
- Cross-reactivity of 24,25-dihydroxyvitamin D metabolites varied widely ( <5% to 643%) in ligand binding assays, with HPLC/UV and LC-MS/MS resolving these metabolites effectively.
Conclusions:
- Significant interference from vitamin D metabolites, particularly 3-epi-25-OHD3 and 24,25-dihydroxyvitamin D, impacts the accuracy of routine 25-OHD assays.
- Method-specific interference varies, with antibody-based and ligand-binding assays showing greater susceptibility compared to HPLC/UV and LC-MS/MS.
- Accurate measurement of 25-OHD requires methods capable of resolving interfering metabolites, highlighting the importance of external quality assessment schemes like DEQAS.
More Related Videos
10:46A Method of Trigonometric Modelling of Seasonal Variation Demonstrated with Multiple Sclerosis Relapse Data
Published on: December 9, 2015
05:03Author Spotlight: Investigating Physiological Functions of Vitamin A Transporters Using HPLC-Based Vitamin A Profiling
Published on: December 27, 2024
Related Concept Videos
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
Role of Skin in Vitamin D Synthesis
The solar UV B rays (290-315 nm) are absorbed by the skin, and 7-dehydrocholesterol (provitamin D3) photolyzes it to previtamin D3, which undergoes a rapid transformation to vitamin...
Factors Affecting Protein-Drug Binding: Drug Interactions
Displacement interactions can have varying outcomes, ranging from toxicity to virtually...