Note: A large open ratio, time, and position sensitive detector for time of flight measurements in UHV
S Lupone1, S Damoy1, A Husseen1
1Institut des Sciences Moléculaires d'Orsay (ISMO), CNRS, Univ. Paris-Sud, Université Paris-Saclay, Orsay F-91405, France.
Abstract:
We report on the construction of an UHV compatible 40 mm active diameter detector based on micro channel plates and assembled directly on the feed-throughs of a DN63CF flange. It is based on the charge division technique and uses a standard 2 inch Si wafer as a collector. The front end electronic is placed directly on the air side of the flange allowing excellent immunity to noise and a very good timing signal with reduced ringing. The important aberrations are corrected empirically providing an absolute positioning accuracy of 500 μm while a 150 μm resolution is measured in the center.