Effect of an arm traction device on image quality and radiation exposure during neck computed tomography
Young Jun Choi1, Jeong Hyun Lee1, Woong Yong Cho1
1Department of Radiology and Research Institute of Radiology, University of Ulsan College of Medicine, Asan Medical Center, 86 Asanbyeongwon-Gil, Songpa-Gu, Seoul 138-736, Republic of Korea.
Objective:
To retrospectively determine the effect of an arm traction device on image quality and radiation exposure during a neck computed tomography (CT) scan.
Materials And Methods:
Standard neck CT examinations with an automatic tube current modulation technique were compared for two groups (intervention group: patients with an arm traction device, n=45; control group: no particular positioning optimization, n=45). Image quality was the primary outcome and was assessed using image noise and the streak artifact. The secondary outcome was radiation exposure, which was measured by the volume CT dose index (CTDIvol) and dose-length product. Potential confounders, including the effective diameter of the neck and scan length, were also assessed.
Results:
Image noise and the streak artifact at the lower neck and the supraclavicular fossa were significantly improved in the intervention group compared with the control group (p<0.001). There was a significant decrease in the CTDIvol in the intervention group versus the control group (p=0.042). DLP showed a tendency toward a decrease in the intervention group that was non-significant (p=0.106). The effective diameter and scan length showed no statistical difference between the two groups.
Conclusion:
An arm traction device improves the image quality in the lower neck and the supraclavicular fossa during a neck CT. Application of this device also reduces the tendency for radiation exposure.
More Related Videos
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Radiological Investigation I: X-ray and CT
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT
X-ray Imaging