Accuracy of a real-time continuous glucose monitoring system in children with septic shock: A pilot study
Sumant Prabhudesai1, Amruta Kanjani2, Isha Bhagat3
1Royal Manchester Children's Hospital, Manchester, UK.
Insights
Continuous glucose monitoring systems (CGMS) offer a fairly accurate estimate of blood glucose in children with septic shock. Accuracy may be a concern at glucose extremes, warranting further study.
Area of Science:
- Pediatric Intensive Care
- Endocrinology
- Medical Devices
Background:
- Septic shock in children presents significant challenges in glucose management.
- Accurate real-time glucose monitoring is crucial for timely intervention.
Purpose of the Study:
- To evaluate the accuracy of a real-time continuous glucose monitoring system (CGMS) in pediatric patients with septic shock.
- To compare interstitial glucose readings from CGMS with laboratory blood glucose measurements.
Main Methods:
- Prospective, observational study including children (30 days to 18 years) with septic shock.
- Simultaneous collection of interstitial glucose (CGMS) and blood glucose (BG) readings.
- Statistical analysis included correlation, Clarke's error grid, and receiver operating characteristics (ROC) curve analysis.
Main Results:
- A total of 235 paired BG-CGMS readings were analyzed from 19 children.
- CGMS showed a correlation coefficient of 0.61 (P < 0.001) with BG, with a median relative absolute difference of 17.29%.
- 94.5% of readings fell within clinically acceptable zones (A and B) on Clarke's error grid. Accuracy was not impacted by clinical variables like edema or vasopressors. ROC analysis indicated good predictive value for hypoglycemia (AUC=0.85) and hyperglycemia (AUC=0.87).
Conclusions:
- Real-time CGMS provides a reasonably accurate estimate of blood glucose in pediatric septic shock.
- CGMS performance is robust across various clinical conditions, but accuracy at glucose extremes requires attention.
- Larger studies are recommended to confirm CGMS utility for detecting hypoglycemia and hyperglycemia.
Aims:
The aim of this prospective, observational study was to determine the accuracy of a real-time continuous glucose monitoring system (CGMS) in children with septic shock.
Subjects And Methods:
Children aged 30 days to 18 years admitted to the Pediatric Intensive Care Unit with septic shock were included. A real-time CGMS sensor was used to obtain interstitial glucose readings. CGMS readings were compared statistically with simultaneous laboratory blood glucose (BG).
Results:
Nineteen children were included, and 235 pairs of BG-CGMS readings were obtained. BG and CGMS had a correlation coefficient of 0.61 (P < 0.001) and a median relative absolute difference of 17.29%. On Clarke's error grid analysis, 222 (94.5%) readings were in the clinically acceptable zones (A and B). When BG was < 70, 70-180, and > 180 mg/dL, 44%, 100%, and 76.9% readings were in zones A and B, respectively (P < 0.001). The accuracy of CGMS was not affected by the presence of edema, acidosis, vasopressors, steroids, or renal replacement therapy. On receiver operating characteristics curve analysis, a CGMS reading <97 mg/dL predicted hypoglycemia (sensitivity 85.2%, specificity 75%, area under the curve [AUC] =0.85). A reading > 141 mg/dL predicted hyperglycemia (sensitivity 84.6%, specificity 89.6%, AUC = 0.87).
Conclusion:
CGMS provides a fairly, accurate estimate of BG in children with septic shock. It is unaffected by a variety of clinical variables. The accuracy over extremes of blood sugar may be a concern. We recommend larger studies to evaluate its use for the early detection of hypoglycemia and hyperglycemia.
Related Concept Videos
Hypoglycemia and Glucagon
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
SBAR II: Application of SBAR
SBAR Report from a Nurse to a Health Care Provider
S: "Hello, Dr. Smith. This is Jane, RN, from the Med Surg unit. I am calling to tell you about Ms. White in Room 210, who is experiencing increased pain and redness at her incision site. Her recent...
Errors occurring during blood pressure monitoring
Several factors...