Related Experiment Video
Updated: Mar 27, 2026

Mass Spectrometric Analysis of Glycosphingolipid Antigens
Published on: April 16, 2013
Analysis and Quantitation of Glycated Hemoglobin by Matrix Assisted Laser Desorption/Ionization Time of Flight Mass
Stephen J Hattan1, Kenneth C Parker2, Marvin L Vestal2
1SimulTOF Systems, Sudbury, MA, 01776, USA. stephen.hattan@simultof.com.
Abstract:
Measurement of glycated hemoglobin is widely used for the diagnosis and monitoring of diabetes mellitus. Matrix assisted laser desorption/ionization (MALDI) time of flight (TOF) mass spectrometry (MS) analysis of patient samples is used to demonstrate a method for quantitation of total glycation on the β-subunit of hemoglobin. The approach is accurate and calibrated with commercially available reference materials. Measurements were linear (R(2) > 0.99) across the clinically relevant range of 4% to 20% glycation with coefficients of variation of ≤ 2.5%. Additional and independent measurements of glycation of the α-subunit of hemoglobin are used to validate β-subunit glycation measurements and distinguish hemoglobin variants. Results obtained by MALDI-TOF MS were compared with those obtained in a clinical laboratory using validated HPLC methodology. MALDI-TOF MS sample preparation was minimal and analysis times were rapid making the method an attractive alternative to methodologies currently in practice.
Related Concept Videos
Matrix-Assisted Laser Desorption Ionization (MALDI)
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...
MALDI-TOF Mass Spectrometry