Urinary Bladder Distention Evoked Visceromotor Responses as a Model for Bladder Pain in Mice
Published on: April 27, 2014
Distributed pressure sensors for a urethral catheter
A new flexible sensor strip for urethral catheters measures pressure during urodynamic testing. This innovation aids in diagnosing urinary incontinence causes with high sensor resolution.
Area of Science:
- Biomedical Engineering
- Medical Devices
- Sensor Technology
Background:
- Urodynamic testing is crucial for diagnosing urinary incontinence.
- Accurate urethral pressure measurement is essential for effective diagnosis.
- Existing methods may have limitations in providing continuous, localized pressure data.
Purpose of the Study:
- To develop a flexible sensor strip for urethral catheters.
- To enable precise measurement of pressure within the human urethra.
- To improve the diagnostic capabilities for urinary incontinence.
Main Methods:
- Fabrication of capacitive pressure sensors on a flexible polyimide-copper substrate.
- Utilizing surface micromachining and precise alignment/assembly techniques.
- Experimental evaluation using an in vitro test rig with a pressure chamber.
Main Results:
- The developed sensor strip demonstrated adequate sensitivity and repeatability.
- Individual sensor calibration factors showed variation.
- The least sensitive sensor achieved a resolution better than 0.1 psi.
Conclusions:
- The instrumented urethral catheter is suitable for clinical urodynamic testing.
- The sensor strip offers a promising tool for diagnosing urinary incontinence.
- The technology provides high-resolution pressure measurements in the urethra.
More Related Videos
09:17In Vivo Luminal Measurement of Distension-Evoked Urothelial ATP Release in Rodents
Published on: September 7, 2022
10:07Cystometric and External Urethral Sphincter Measurements in Awake Rats with Implanted Catheter and Electrodes Allowing for Repeated Measurements
Published on: January 30, 2018
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Pressure Gauges
Measurement of Fluid Pressure
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...
Urodynamic Studies: Uroflowmetry
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment: