Related Experiment Video
Updated: Mar 27, 2026

Multi-Modal Home Sleep Monitoring in Older Adults
Published on: January 26, 2019
[Comparative study on electronic sphygmomanometer and mercury sphygmomanometer among adults in the field study]
Objective:
To compare the electronic sphygmomanometer with mercury sphygmomanometer on measuring the average blood pressure, prevalence of hypertension in a same population. To set up a numerical relationship between two blood pressure value of the two devices.
Methods:
544 adults of ≥ 18 years old were from 6 neighborhood of Hunan Province. OMRON HEM-907 and mercury sphygmomanometer were used to test the difference. The blood pressure value of mercury sphygmomanometer was gold standard. The difference of systolic pressure and diastolic pressure of two devices was tested using paired t test and Bland-Altman analysis. Chi-square test was used to analyze the two prevalence of hypertension among the same population. The equation of linear regression was established using the value of blood pressure of two devices.
Results:
Paired t test showed the statistical difference on the systolic pressure and diastolic pressure of two devices (P < 0.0001). Chi-square test presented a higher prevalence of hypertension by using the electronic sphygmomanometer (P < 0.0001). Bland-Altman analysis showed mercury sphygmomanometer couldn't be simply replaced by electronic sphygmomanometer. The two independent equation of linear regression was calculated on systolic pressure and diastolic pressure.
Conclusion:
The difference still existed when using the electronic sphygmomanometer and mercury sphygmomanometer at a large scale population. It's important to conduct the comparative study on electronic sphygmomanometer and mercury sphygmomanometer.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(two-step method)
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Measurement of Blood Pressure