Suture slippage in knotless suture anchors resulting in subacromial-subdeltoid bursitis
Mohammad Reza Hayeri1,2, Daniel T Keefe3, Eric Y Chang4,5
1Radiology Service, VA San Diego Healthcare System, 3350 La Jolla Village Drive, MC 114, San Diego, CA, 92161, USA.
Abstract:
Rotator cuff repair using a suture bridge and knotless suture anchors is a relatively new, but increasingly used technique. The suture bridge technique creates an anatomically similar and more secure rotator cuff repair compared with conventional arthroscopic techniques and the use of knotless anchors eliminates the challenges associated with knot tying during arthroscopic surgery. However, previous in vitro biomechanical tests have shown that the hold of the suture in a knotless suture anchor is far lower than the pullout strength of the anchor from bone. Up until now slippage has been a theoretical concern. We present a prospectively diagnosed case of in vivo suture loosening after rotator cuff repair using a knotless bridge technique resulting in subacromial-subdeltoid bursitis.