Acoustic radiation force impulse with shear wave speed quantification of pancreatic masses: A prospective study
Mirko D'Onofrio1, Riccardo De Robertis1, Stefano Crosara1
1Department of Radiology, G.B. Rossi Hospital, University of Verona, Piazzale L.A. Scuro 10, 37134 Verona, Italy.
Background:
Acoustic Radiation Force Impulse (ARFI) is a new ultrasound technique that evaluates mechanical properties of tissues. To evaluate the use of ARFI with shear waves speed quantification for pancreatic masses characterization during the ultrasound examination.
Methods:
123 pancreatic lesions were prospectively evaluated. Median shear waves speeds were compared with Mann-Whitney U test. Two reading methods were applied for the characterization of adenocarcinoma: more than one measurement above the top shear waves speed (SWS) value. Two reading methods were applied to diagnose mucinous lesion: at least 2 (method 1) or 3 (method 2) numerical measurements. Sensitivity, specificity, positive and negative predictive values and accuracy of each reading method were calculated. Forty volunteers were included for normal ARFI values.
Results:
In the adenocarcinoma group median SWS value was 2.74 m/s. In the volunteers group the median SWS value was 1.17 m/s. Significant difference between SWS median values of adenocarcinoma and normal pancreas was found (P < 0.05). For the diagnosis of pancreatic solid masses if more than one measurement is above the top SWS value of 4.00 m/s results in the study, the diagnosis of ductal adenocarcinoma is highly specific with specificity and PPV of 100%. Good sensitivity (73.3%) and specificity (83.3%) were obtained for the characterization of mucinous cystic lesions.
Conclusions:
Acoustic Radiation Force Impulse imaging could help in the non-invasive characterization of solid and cystic lesions of the pancreas during a conventional US examination.
More Related Videos
05:56Implementation of Non-invasive Point of Care Transient Elastography for Evaluation of Liver Disease in Pediatric Populations with Cystic Fibrosis
Published on: August 29, 2025
12:18Magnetic Resonance Elastography Methodology for the Evaluation of Tissue Engineered Construct Growth
Published on: February 9, 2012
Related Concept Videos
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Assessment of the Abdomen II: Percussion
Percussion
Percussion is an essential...