Coronary CT angiography derived fractional flow reserve: Methodology and evaluation of a point of care algorithm
Adriaan Coenen1, Marisa M Lubbers1, Akira Kurata2
1Department of Radiology, Erasmus University Medical Center, Rotterdam, The Netherlands; Department of Cardiology, Erasmus University Medical Center, Rotterdam, The Netherlands.
Background:
Recently several publications described the diagnostic value of coronary CT angiography (coronary CTA) derived fractional flow reserve (CTA-FFR). For a recently introduced on-site CTA-FFR application, detailed methodology and factors potentially affecting performance have not yet been described.
Objective:
To provide a methodological background for an on-site CTA-FFR application and evaluate the effect of patient and acquisition characteristics.
Methods:
The on-site CTA-FFR application utilized a reduced-order hybrid model applying pressure drop models within stenotic regions. In 116 patients and 203 vessels the diagnostic performance of CTA-FFR was investigated using invasive FFR measurements as a reference. The effect of several potentially relevant factors on CTA-FFR was investigated.
Results:
90 vessels (44%) had a hemodynamically relevant stenosis according to invasive FFR (threshold ≤0.80). The overall vessel-based sensitivity, specificity and accuracy of CTA-FFR were 88% (CI 95%:79-94%), 65% (55-73%) and 75% (69-81%). The specificity was significantly lower in the presence of misalignment artifacts (25%, CI: 6-57%). A non-significant reduction in specificity from 74% (60-85%) to 48% (26-70%) was found for higher coronary artery calcium scores. Left ventricular mass, diabetes mellitus and large vessel size increased the discrepancy between invasive FFR and CTA-FFR values.
Conclusions:
On-site calculation of CTA-FFR can identify hemodynamically significant CAD with an overall per-vessel accuracy of 75% in comparison to invasive FFR. The diagnostic performance of CTA-FFR is negatively affected by misalignment artifacts. CTA-FFR is potentially affected by left ventricular mass, diabetes mellitus and vessel size.
More Related Videos
05:07Author Spotlight: Improved Localization and Monitoring of Coronary Flow Reserve Using Modified PLAX View in Mice
Published on: August 25, 2023
06:39Ultrasound Based Assessment of Coronary Artery Flow and Coronary Flow Reserve Using the Pressure Overload Model in Mice
Published on: April 13, 2015
Related Concept Videos
Acute Coronary Syndrome III: Diagnostic Studies
Imaging Studies for Cardiovascular System V: CT