Comparison of CAIPIRINHA-VIBE, Radial-VIBE, and conventional VIBE sequences for dynamic contrast-enhanced (DCE) MRI:
Bohyun Kim1, Chang Kyung Lee2, Nieun Seo3
1Department of Radiology and Research Institute of Radiology, Asan Medical Center, University of Ulsan College of Medicine, 88 Olympic-ro, Seoul, Republic of Korea; Department of Radiology, Ajou University School of Medicine, 165 Worldcup-ro, Suwon, Republic of Korea.
Objective:
To validate radial acquisition of volumetric interpolated breath hold examination (Radial-VIBE) and the controlled aliasing in parallel imaging results in higher acceleration (CAIPIRINHA-VIBE) sequences for dynamic contrast-enhanced MRI (DCE-MRI) by comparing them to conventional VIBE sequence using a phantom.
Methods:
On a DCE-MRI phantom containing various concentrations of NiCl2 solutions, six minutes of dynamic series and T1 mapping with variable flip angle methods were acquired using conventional VIBE, Radial-VIBE, and CAIPIRINHA-VIBE sequences on 3.0-T scanners. Signal stability and signal linearity were tested for dynamic series and the precision of R1 values were tested for T1 mapping series. The scans were repeatedly performed at two weeks and three months to test repeatability/reproducibility, assessed by within-subject coefficient of variation (WSCV).
Results:
Signal stability over six minutes was excellent in all three sequences. Regarding the signal linearity, CAIPIRINHA-VIBE demonstrated the highest linear correlation (r=0.963), followed by conventional VIBE (r=0.959) and Radial-VIBE (r=0.953). Regarding the R1 precision, CAIPIRINHA-VIBE (r=0.985) was the most accurate, followed by conventional VIBE (r=0.861) and Radial-VIBE (r=0.442). CAIPIRINHA-VIBE showed excellent repeatability/reproducibility (WSCV, 1.79-6.71%) compared with Radial-VIBE (WSCV, 2.04-67.2%) and conventional VIBE (WSCV, 3.4-31.9%).
Conclusion:
In terms of signal stability, signal linearity, R1 precision, and repeatability/reproducibility, CAIPIRINHA-VIBE demonstrated outstanding performance for DCE-MRI compared with Radial-VIBE and conventional VIBE.
More Related Videos
11:27Preparation and In Vitro Characterization of Dendrimer-based Contrast Agents for Magnetic Resonance Imaging
Published on: December 4, 2016
15:48Tracking the Mammary Architectural Features and Detecting Breast Cancer with Magnetic Resonance Diffusion Tensor Imaging
Published on: December 15, 2014
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies IV: Magnetic Resonance Imaging
Imaging Studies for Cardiovascular System IV: CMRI
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
NMR Spectrometers: Radiofrequency Pulses and Pulse Sequences