Related Experiment Video
Updated: Mar 27, 2026

Treatment with Locking Intramedullary Nailing for Intertrochanteric Fracture of the Femur Utilizing a New Awl with a Distal Positioner
Published on: June 6, 2025
[BIOMECHANICAL RESEARCH OF DIFFERENT INTERNAL FIXATIONS USING LOCKING RECONSTRUCTION PLATE FOR ACETABULAR TRANSVERSE
Objective:
To compare the biomechanical stability of acetabular transverse fracture fixed with single column locking reconstruction plate, locking reconstruction plate combined with lag screw, and double columns lag screws.
Methods:
Acetabular transverse fractures were established in 20 adult fresh semi-pelvis specimens, and divided into 5 groups randomly (n=4). The anterior columns were fixed with locking reconstruction plates in group A; the posterior columns were fixed with locking reconstruction plates in group B; the anterior columns were fixed with locking reconstruction plates and the posterior columns with lag screws in group C; the anterior columns were fixed with lag screws and the posterior columns with locking reconstruction plates in group D; the double columns were fixed with lag screws in group E. When loading three times weight, the longitudinal displacement of fracture fragment was measured, and the shear rigidity was calculated to compare the stability among groups.
Results:
The longitudinal displacement of groups A and B were significantly greater than that of groups C, D, and E (P<0.05), and the shear rigidity of groups A and B were significantly lower than that of groups C, D, and E (P< 0.05). The longitudinal displacement of group B was significantly greater than that of group A (P< 0.05), but no significant difference was found in the shear rigidity between groups A and B (P>0.05). There was no significant difference in the longitudinal displacement and shear rigiditramong groups C, D, and E (P>0.05).
Conclusion:
Locking reconstruction plate combined with lag screw and double columns lag screws have similar stability, they have stronger stability than the single column locking reconstruction plate. The stability of anterior column locking reconstruction plate is better than that of the posterior column locking reconstruction plate.
More Related Videos
05:34A Mini-Invasive Internal Fixation Technique for Studying Immobilization-Induced Knee Flexion Contracture in Rats
Published on: May 20, 2019
15:11Surgical Fixation of Sternal Fractures: Preoperative Planning and a Safe Surgical Technique Using Locked Titanium Plates and Depth Limited Drilling
Published on: January 5, 2015