Related Experiment Video
Updated: Mar 27, 2026

ELIME Enzyme Linked Immuno Magnetic Electrochemical Method for Mycotoxin Detection
Published on: October 23, 2009
Asymmetric Mach-Zehnder Interferometer Based Biosensors for Aflatoxin M1 Detection
Tatevik Chalyan1, Romain Guider2, Laura Pasquardini3
1Nanoscience Laboratory, Department of Physics, University of Trento, Via Sommarive 14, 38123 Povo (TN), Italy. tatevik.chalyan@unitn.it.
Abstract:
In this work, we present a study of Aflatoxin M1 detection by photonic biosensors based on Si₃N₄ Asymmetric Mach-Zehnder Interferometer (aMZI) functionalized with antibodies fragments (Fab'). We measured a best volumetric sensitivity of 10⁴ rad/RIU, leading to a Limit of Detection below 5 × 10(-7) RIU. On sensors functionalized with Fab', we performed specific and non-specific sensing measurements at various toxin concentrations. Reproducibility of the measurements and re-usability of the sensor were also investigated.