Comparison of ActiGraph GT3X+ and StepWatch Step Count Accuracy in Geriatric Rehabilitation Patients
Sandra C Webber1, Philip D St John
1Department of Physical Therapy, College of Rehabilitation Sciences, University of Manitoba, Winnipeg, MB, Canada.
Abstract:
Activity monitors may not accurately detect steps in hospitalized older adults who walk slowly. We compared ActiGraph GT3X+ step counts (hip and ankle locations, default and low frequency extension [LFE] analyses) to the StepWatch monitor (ankle) during a hallway walk in 38 geriatric rehabilitation patients (83.2 ± 7.1 years of age, 0.4 ± 0.2 m/s gait speed). Absolute percent error values were low (<3%) and did not differ for the StepWatch and the GT3X+ (ankle, LFE); however, error values were high (19-97%) when the GT3X+ was worn at the hip and/ or analyzed with the default filter. Although these finding suggest the GT3X+ (ankle, LFE) functions as well as the StepWatch in detecting steps during walking in older adults with slow gait speeds, further research is needed to determine whether the GT3X+ is also able to disregard other body movements (e.g., fidgeting) that occur when full day monitoring is utilized.