Performance comparison of a new automated cuff pressure controller with currently available devices in both basic
Junichi Michikoshi1, Shigekiyo Matsumoto2, Hiroshi Miyawaki3
1Department of Anesthesiology and Intensive Care, Faculty of Medicine, Oita University, 1-1 Idaigaoka-Hasamamachi, Yufu City, Oita 879-5593 Japan ; Department of Medical Technologists, Division of Engineering, Kokura Memorial Hospital, 3-2-1 Asano-Kokura Kitaku, Kitakyushu City, Fukuoka 802-8555 Japan.
Background:
The management of tracheal tube cuff pressure in patients receiving mechanical ventilation is important for the prevention of ventilator-associated pneumonia. Currently, cuff pressure is intermittently monitored with a pressure gauge and adjusted when necessary in a routine practice. However, this method results in wide variations in pressure, and adequate management is difficult due to the spontaneous release of air from the cuff, which reduces cuff pressure. In order to continuously maintain a uniform cuff pressure, we developed a new automated cuff pressure controller and compared its properties with existing devices.
Methods:
The effectiveness of the new device was assessed with a model trachea/lung and tracheal tube by measuring cuff pressure while on mechanical ventilation. An electrically powered automatic cuff controller or manual cuff pressure control was used for comparison purposes. The effectiveness of the new device was also examined in patients receiving mechanical ventilation by continuously measuring cuff pressure for a 24-h period.
Results:
Cuff pressure was uniformly maintained with the new device. Moreover, in the clinical setting, variation in pressure from the set pressure was minimal with both the new device and existing device, relative to the intermittent monitoring method. This suggests that, as with the existing device, uniform cuff pressure management is possible with the new device.
Conclusions:
Our results demonstrate the ability of the new cuff pressure controller to manage cuff pressure without the need of a power source, highlighting its potential utility in clinical settings.
Related Concept Videos
Basic Equation for Pressure Field
Energy Basics
The Sense of Self: Reflected Self-Appraisal and Social Comparison
VSEPR Theory and the Basic Shapes
Basicity of Aliphatic Amines
To measure the basicity of amines, two conventions are generally used. The first defines Kb as the basicity constant for the deprotonation reaction of water by the amine, as presented in Figure 1. Conventionally, lower Kb indicates higher...
Basic Operations on Signals
Time Reversal mirrors a continuous-time signal about the vertical axis at t=0. This is achieved by substituting t with −t. For example, if a signal x(t) is considered, the time-reversed signal is x(−t). This operation can be graphically represented, showing the mirrored signal.