Concentration Monitoring of Plasma Ribavirin: Validation of a Liquid Chromatography-Mass Spectrometric Method and
Nigel W Brown1, Phillip E Morgan, Kosh Agarwal
1Institute of Liver Studies, Denmark Hill, King's College Hospital, London, United Kingdom.
A new LC-MS/MS method accurately measures ribavirin in plasma. Proper sample handling, including refrigeration and prompt separation, is crucial for reliable ribavirin concentration results.
Area of Science:
- Analytical Chemistry
- Pharmacology
Background:
- Development of a liquid chromatography-tandem mass spectrometric (LC-MS/MS) method for routine ribavirin measurement in plasma.
- Addressing the need for precise quantification of antiviral drug concentrations.
Purpose of the Study:
- To establish and validate a robust LC-MS/MS assay for quantifying ribavirin in ethylenediaminetetraacetic acid (EDTA)-anticoagulated plasma.
- To investigate the stability of ribavirin in plasma under various storage conditions.
- To apply the developed method for therapeutic drug monitoring in hepatitis C patients.
Main Methods:
- Utilized a hydrophilic interaction chromatography column with a specific mobile phase (0.1 mmol/L ammonium formate pH 3.0) and acetonitrile gradient.
- Employed tandem mass spectrometry for quantitation using specific ion transitions (m/z 245.3 > 113.0 and m/z 245.3 > 96.0).
- Validated the method for linearity, sensitivity (limit of detection and quantitation), imprecision, and bias, and assessed stability in clinical samples.
Main Results:
- Demonstrated a linear relationship up to 10 mg/L with a limit of quantitation of 0.1 mg/L.
- Achieved assay imprecision <10% at 0.1 mg/L and <6% for concentrations >0.4 mg/L, with bias <4%.
- Found ribavirin stable at room temperature for 3 days, but decreased significantly by day 7; stable for 15 weeks at -20 °C. Plasma concentrations showed instability after separation from whole blood at room temperature.
Conclusions:
- A validated hydrophilic interaction LC-MS/MS method is suitable for measuring ribavirin in plasma.
- Recommended optimal sample handling: keep samples at 4 °C, separate within 2 hours, and store at 4 °C for short-term or -20 °C for long-term analysis.
- Successfully applied the method to monitor ribavirin therapy in hepatitis C patients, observing steady-state concentrations after prolonged treatment.