Effectiveness of Disposable (Single-use) Tips for Dental Air-water Syringes
Background:
Single-use dental syringe (SUS) tips eliminate cross-infection risks. Many dental materials are moisture sensitive; therefore bonding could be affected by remaining moisture from SUS tips.
Methods:
An air-water syringe was calibrated to 50 PSI and the air pressure measured. SUS tips were then attached and pressure re-measured. Seven tips were tested; moisture content and spray pattern were investigated. Ethical approval was granted. Twenty third molar teeth were sectioned, stored in phosphate buffered saline and randomly allocated into groups with a SUS tip. They were mounted in acrylic and polished. Specimens were dried with a tip under standardised conditions, treated using a 3-step etch-prime-bond technique, then had a post bonded perpendicular to the dentine surface. The force to detach them was measured. Data were analysed using one-way analysis of variance (ANOVA) and Chi-squared tests.
Results:
The difference in output pressure between groups proved significant (p < 0.001). Tensile force at failure was significantly different between groups (p = 0.19).
Conclusions:
The pressure output of the syringe and bond strength are influenced by the SUS tip.
Related Concept Videos
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils,...
Suctioning the Nasopharyngeal Airway
Equipment Required
Suctioning the Oropharyngeal Airway
After assembling the equipment, the nurse should practice hand hygiene and don appropriate PPE according to infection control guidelines to avoid the...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...