Sensory function assessment of the human male lower urinary tract using current perception thresholds
Stephanie C Knüpfer1, Martina D Liechti1, Flavia Gregorini1
1Department of Neuro-Urology, Spinal Cord Injury Center and Research, Balgrist University Hospital, University of Zürich, Zürich, Switzerland.
Current perception threshold (CPT) measurements are reliable for assessing sensation in distinct male lower urinary tract (LUT) locations. This method offers valuable insights into afferent LUT nerve sensitivity, complementing standard urodynamic tests.
Area of Science:
- Urology
- Neuroscience
- Sensory Physiology
Background:
- Standard urodynamic investigations do not fully assess afferent nerve sensitivity in the lower urinary tract (LUT).
- Current perception threshold (CPT) measurement offers a potential method for evaluating sensory function in the LUT.
Purpose of the Study:
- To determine the feasibility and reliability of CPT measurements for sensory assessment at various locations within the male LUT.
- To explore the potential of CPT as a complementary diagnostic tool for LUT afferent nerve function.
Main Methods:
- Twelve healthy adult males underwent repeated CPT measurements at the bladder dome, trigone, and proximal, membranous, and distal urethra.
- Electrical stimulation was applied transurethrally using specific frequencies and pulse durations, with bladder volume maintained consistently.
- Analysis of variance (ANOVA) and intraclass correlation coefficient (ICC) were used to assess measurement reliability.
Main Results:
- CPT varied significantly based on stimulation site and type, with lower thresholds observed for 0.5 Hz/1 ms stimulation compared to 3 Hz/0.2 ms.
- CPTs were higher at the bladder dome than in the proximal and distal urethra.
- Reliability was generally excellent to good (ICC 0.67-0.96), with exceptions noted for the bladder dome and distal urethra under specific stimulation conditions.
Conclusions:
- CPT measurements are feasible and reliable for assessing sensation in most male LUT locations.
- Alert and compliant participants are crucial for accurate CPT assessment.
- CPT measurement can serve as a valuable complementary method to assess afferent LUT nerve sensitivity, particularly for urethral afferents.
Related Concept Videos
Anatomy of the Genitourinary System II: Bladder and Urethra
Nursing Assessment of the Genitourinary System III: Percussion and Auscultation
Urodynamic Studies: Uroflowmetry
Urine Studies II: Urine Culture and Sensitivity Test
Nursing Assessment of the Genitourinary System II: Inspection and Palpation
Nursing Assessment of the Genitourinary System I: Health History