Purpose:
To evaluate obtained data by using a relatively novel devices and their results which are important eg. in refractive and cataract surgery.
Material And Methods:
The study included 66 eyes (n = 66). Subjects were represented by 32 women and 1 man whose age was 22.5 years ± 1.2 years (min. 21, max. 26 years) without any signs of potential eye disease. Duration of the study was 3 months. Results were compared with the measurements using the auto-refract-keratro-tono-pachymeter (TRK 1P, Topcon, Japan), Keratograph 5M (Oculus, Germany) and aberrometer iTRACE (Hoya, Japan).
Results:
After 3 months were statistically compared keratometry values of corneal anterior surface detected by all devices. They provided to be comparable. Furthermore the values of objective refraction and pachymetry were detected.
Conclusion:
Results of this study show a statistically significant correlation values of objective refraction using devices TRK and iTRACE (r = 0.66 at p = 0.05) and showed a significant relationship between the keratometric data for all the devices. All used methods and devices are possible to reliably and use for valid evaluation parameters of the eye.
Key Words:
aberrometry, low and high orders abberations, keratometry, topography, pachymetry.
More Related Videos
07:14Tracking Infiltration Front Depth Using Time-lapse Multi-offset Gathers Collected with Array Antenna Ground Penetrating Radar
Published on: May 1, 2018
10:45Real-time Iontophoresis with Tetramethylammonium to Quantify Volume Fraction and Tortuosity of Brain Extracellular Space
Published on: July 24, 2017
Related Concept Videos
Attenuated Total Reflectance (ATR) Infrared Spectroscopy: Overview
The ATR process begins by directing a beam...
Distance Measurements by Taping
Precipitation Gravimetry
In determining nickel by gravimetric analysis, a precipitant of ethanolic dimethylglyoxime is added to a hot nickel salt solution. This is quickly followed by the dropwise addition of dilute ammonia solution until precipitation occurs. A...