Multiple calibrator measurements improve accuracy and stability estimates of automated assays
Neval Akbas1, Jeffrey R Budd2, George G Klee1
1a Department of Laboratory Medicine and Pathology , Mayo Clinic , Rochester , MN , USA ;
Objectives:
The effects of combining multiple calibrations on assay accuracy (bias) and measurement of calibration stability were investigated for total triiodothyronine (TT3), vitamin B12 and luteinizing hormone (LH) using Beckman Coulter's Access 2 analyzer.
Methods:
Three calibration procedures (CC1, CC2 and CC3) combined 12, 34 and 56 calibrator measurements over 1, 2, and 3 days. Bias was calculated between target values and average measured value over 3 consecutive days after calibration. Using regression analysis of calibrator measurements versus measurement date, calibration stability was determined as the maximum number of days before a calibrator measurement exceeded 5% tolerance limits.
Results:
Competitive assays (TT3, vitamin B12) had positive time regression slopes, while sandwich assay (LH) had a negative slope. Bias values for TT3 were -2.49%, 1.49%, and -0.50% using CC1, CC2 and CC3 respectively, with calibrator stability of 32, 20, and 30 days. Bias values for vitamin B12 were 2.44%, 0.91%, and -0.50%, with calibrator stability of 4, 9, and 12 days. Bias values for LH were 2.26%, 1.44% and -0.29% with calibrator stability of >43, 39 and 36 days. Measured stability was more consistent across calibration procedures using percent change rather than difference from target: 26 days for TT3, 12 days for B12 and 31 days for LH.
Conclusions:
Averaging over multiple calibrations produced smaller bias, consistent with improved accuracy. Time regression slopes in percent change were unaffected by number of calibration measurements but calibrator stability measured from the target value was highly affected by the calibrator value at time zero.
Related Concept Videos
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Glassware Calibration
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
Uncertainty in Measurement: Accuracy and Precision
Systematic Error: Methodological and Sampling Errors
Sampling errors originate from improper sampling methods or the wrong sample population. These errors can be minimized by refining the sampling strategy. Defective instruments or faulty calibrations are the sources of instrumental...
Uncertainty in Measurement: Reading Instruments
Calibration Curves: Linear Least Squares
For data that follow a straight line, the standard method for fitting is the linear...