Related Experiment Video
Updated: Mar 26, 2026

Flow Cytometry to Estimate Leukemia Stem Cells in Primary Acute Myeloid Leukemia and in Patient-derived-xenografts, at Diagnosis and Follow Up
Published on: March 26, 2018
Decreased sensitivity of the automatic white precursor cell channel (WPC) for blast flagging in patients with
Nuri Lee1, Ju Hyun Jun1, Dong Soon Lee1
1Department of Laboratory Medicine, Seoul National University College of Medicine, Seoul, South Korea.
Objectives:
The advantage of Sysmex XN system is its performance of an automatic reflex test in white cell precursor (WPC) channel, which gives an accurate differential count. We performed a real-time evaluation of the automatic differential count according to WBC number with Sysmex XN series and demonstrated a significant differential impact on blast distinction depending on the number of WBC.
Design And Methods:
We categorized the 49,699 specimen according to WBC number and compared the results of blast flagging in the white cell differential (WDF) channel and WPC channel on the basis of the results of manual differential count. Additionally, clinical impact of missed blasts after running WPC reflex test was analyzed.
Results:
For patients with WBC under 1.5×10(9)/L, blast flagging for WPC channel showed markedly decreased sensitivity (56%) compared to that of WDF mode (100%). Most of the patients with missed blasts in WPC mode were under chemotherapy or hematopoietic stem cell transplantation. The specificity and efficiency for WPC channel were much higher than that of WDF mode in all range of WBC.
Conclusions:
Considering the additional reagent cost required for the WPC channel, WPC reflex test is not suitable for patients with leukopenia if it is operated alone. Instead, manual blood film review or double-check with other supplemental equipment should be accompanied.
More Related Videos
09:57Comprehensive Protocol to Sample and Process Bone Marrow for Measuring Measurable Residual Disease and Leukemic Stem Cells in Acute Myeloid Leukemia
Published on: March 5, 2018
05:58Evaluation of a Point-of-Care Testing Analyzer for Measuring Peripheral Blood Leukocytes
Published on: March 22, 2022
Related Concept Videos
Disorders of Leukocytes
Leukopenia may result from bone marrow disorders, autoimmune diseases, and infectious diseases. For example, conditions such as multiple myeloma and aplastic anemia can impair the bone marrow's ability to produce adequate leukocytes. Similarly, autoimmune diseases like lupus and viral infections such as HIV can prompt the immune...
Structure and Function of Leukocytes
White blood cells protect the body...