Comparison of proximal humeral bone stresses between stemless, short stem, and standard stem length: a finite element
Najmeh Razfar1, Jacob M Reeves2, Daniel G Langohr1
1Department of Biomedical Engineering, Western University, London, ON, Canada; Roth|McFarlane Hand and Upper Limb Centre, St. Joseph's Health Centre, London, ON, Canada; Department of Mechanical Engineering, Western University, London, ON, Canada.
Background:
The stem lengths of humeral components used in shoulder arthroplasty vary; however, the literature on these devices is limited. This finite element study investigates the effect of humeral component length on stresses in the proximal humerus.
Methods:
Intact and 3 reconstructed (standard length, short, and stemless implants) finite element models were created from shoulder computed tomography scan data (N = 5). Loading was simulated at varying abduction angles (15°, 45°, and 75°). The average bone stress (represented as a percentage of intact values) was reported at 8 transverse slices. In addition, the overall average change in cortical and trabecular bone stresses was quantified.
Results:
Cortical bone stresses in the most proximal slice for the standard (58% ± 12%) and short (78% ± 10%) stem models were significantly reduced compared with the intact (100%) and stemless (101% ± 6%) models (P = .005). These reductions persisted in the second cortical slice for the standard stem compared with the intact, stemless, and short models (P = .025). Interestingly, stresses in the trabecular bone within these proximal slices were significantly elevated when stemless implants were used compared with all other implants (P < .001), regardless of abduction angle.
Conclusion:
Reducing stem length produced humeral stresses that more closely matched the intact stress distribution in proximal cortical bone. Opposing trends presented in the proximal trabecular bone, probably because of differences in load transfer when shorter stems are used. Accordingly, the results suggest that implant stem length is 1 variable that can be modified in an attempt to better mimic intact bone stresses during humeral component insertion, provided stem fixation is adequate.
Related Concept Videos
Applications of Stress
The...
Stresses under Combined Loadings
The process begins by slicing the tube at critical points and analyzing the internal forces and stress components at these sections, focusing on the centroid. Normal stresses, generated by axial forces and bending moments, are either compressive or tensile and vary across the section from...
Flexural Stress
Hooke's Law states that within the material's elastic limits, stress is directly proportional to strain. In a member experiencing a bending moment, the strain at any point is relative to its distance...
Normal Strain under Axial Loading
Generalized Hooke's Law
Bending of Members Made of Several Materials
Hooke's Law determines stress in each material, stating that stress is proportional to strain but varies due to each material's...