Increased Computed Tomography Dose Due to Miscentering With Use of Automated Tube Voltage Selection: Phantom and
Peter D Filev1, Pardeep K Mittal1, Xiangyang Tang1
1Department of Radiology and Imaging Sciences, Emory University School of Medicine, Atlanta, GA.
Abstract:
The purpose of the article is to determine if miscentering affected dose with use of automated tube voltage selection software. An anthropomorphic phantom was imaged at different table heights (centered in the computed tomography [CT] gantry, and -6, -3, +3, and +5.7cm relative to the centered position). Topogram magnification, tube voltage selection, and dose were assessed. Effect of table height on dose also was assessed retrospectively in human subjects (n = 50). When the CT table was positioned closer to the x-ray source, subjects appeared up to 33% magnified in topogram images. When subjects appeared magnified in topogram images, automated software selected higher tube potentials and tube currents that were based on the magnified size of the subject rather than the subject׳s true size. Table height strongly correlated with CT dose index (r = 0.98, P < 0.05) and dose length product (r = 0.98, P < 0.05) in the phantom study. Transverse dimension in the topogram highly correlated with dose in human subjects (r = 0.75-0.87, P <0.05). Miscentering results in increased dose due to topogram magnification with automated voltage selection software.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT