Patient Vertical Centering and Correlation with Radiation Output in Adult Abdominopelvic CT
1Department of Radiology, Keck School of Medicine of USC, Los Angeles, CA, USA. phillip.cheng@med.usc.edu.
Abstract:
The purpose of this study was to determine if there is a significant effect, independent of patient size, of patient vertical centering on the current-modulated CT scanner radiation output in adult abdominopelvic CT. A phantom was used to evaluate calculation of vertical positioning and effective diameter at five different table heights. In addition, 656 consecutive contrast-enhanced abdominopelvic scans using the same protocol and automatic tube current modulation settings on a Philips Brilliance 64 MDCT scanner were retrospectively evaluated. The vertical position of the patient center of mass and the average effective diameter of the scanned patient were computed using the reconstructed images. The average volume CT dose index (CTDIvol) for each scan was recorded. The mean patient center of mass y coordinate ranged from -3.7 to 6.7 cm (mean ± SD, 2.8 ± 1.2 cm), indicating that patients were on average positioned slightly below the scanner isocenter. There was a slight tendency for smaller patients to be mis-centered lower than larger patients. Average CTDIvol closely fit a quadratic regression curve with respect to mean effective diameter. However, the value of the regression coefficient relating CTDIvol to the patient's vertical position was nearly zero, indicating only a very slight increase in CTDIvol with patient mis-centering for the scanner used in this study. The techniques used here may be useful both for automated evaluation of proper patient positioning in CT and for estimating the radiation dose effects of patient mis-centering for any CT scanner.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System V: CT
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...