Related Experiment Video
Updated: Mar 26, 2026

A Novel Digital Platform for a Monitored Home-based Cardiac Rehabilitation Program
Published on: April 19, 2019
[Experiences of applying home monitoring for cardiac implantable electronic devices]
Liang Qiu1, Meixiang Xiang, Jian'an Wang
1Department of Cardiology, Second Affiliated Hospital, Zhejiang University College of Medicine, Hangzhou 310009, China.
Objective:
To observe the results of remote home monitoring for cardiac implantable electronic devices (CIED).
Methods:
From November 2009 to July 2014, 37 patients implanted with home monitoring system CIED were enrolled, and 31 cases were implanted with dual chamber pacemakers, 2 with cardiac resynchronization therapy(CRT) and 4 with implantable cardioverter defibrillator (ICD). All patients received clinical routine follow-up.Data was automatic transmitted daily by remote home monitoring system, while special alarm events were transmitted at any time.The success rate of data transmission, abnormal events monitoring and pacing parameter changes were analyzed.
Results:
A total of 37 patients (24 males, mean age (70.0±9.6) years) were enrolled.Average follow-up time was (902±404) days.Success rate of data transmission was 97.6% (32 574), failed data transmission rate was 2.4% (801). Ninth-nine alerts were transmitted from 28 patients (75.7%), including 80 (80.8%) disease-related alerts and 19 (19.2%) system-related alerts.Average detection time of atrial fibrillation by remote home monitoring system to last clinical routine follow-up was 62 (19-120) days.There was no significant change in ventricular threshold, atrial and ventricle sensing during acute, subacute and chronic phases post the application of the steroid-eluting leads(all P>0.05).
Conclusions:
The success rate of data transmission for CIED with remote home monitoring system is efficient, abnormal events can be timely detected.Meanwhile, remote home monitoring system also verified the stability of pacing parameters during acute, subacute and chronic phases post the application of the steroid-eluting leads.
More Related Videos
16:40A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
06:29Wireless Telemetry Device Implantation in a Fontan Ovine Model for Continuous and Long-Term Hemodynamic Monitoring
Published on: May 2, 2025
Related Concept Videos
Holter Monitor: 24-Hour Monitoring
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Cardiomyopathy V: Interprofessional Care
Cardiac Catheterization IV: Nursing Management
Dysrhythmias V: Evaluating Dysrhythmias