Contrast enhanced ultrasound (CEUS) reliably detects critical perfusion changes in compartmental muscle: a model in
R M Sellei1,2, A Waehling3, C D Weber3
1Department of Orthopaedic Trauma at Aachen University, Aachen, Germany. richard.sellei@sana.de.
Purpose:
The purpose of this study was to assess the utility of contrast enhanced ultrasound (CEUS) in the differentiation between physiological and simulated pathophysiological lower limb muscle perfusion pressures in healthy volunteers.
Methods:
The lower limb muscle perfusion pressures in eight healthy volunteers were assessed in the supine position (as a control) and then subsequently in an elevated position with a thigh tourniquet applied to induce venous stasis. An intravenous bolus injection of 2.5 ml contrast agent was given to create a perfusion signal, which was measured with a multiple-frequency probe. Semiquantitative analysis was performed using specific software to create a perfusion curve which allowed measurement of six parameters: the time to arrival (TTA) starting from bolus application (s); peak of signal intensity (%); time to peak (TTP) maximum (seconds); regional blood volume (RBV), regional blood flow (RBF), and mean transit time (MTT) in seconds. Statistical analysis was performed using the Mann-Whitney U test as a non-parametric test (IBM SPSS statistics, version 21, USA).
Results:
The group of simulated hypoperfusion showed significant higher values for TTA (39.8 ± 5.1 s) (p = 0.028), TTP (43.8 ± 13.6 s) (p = 0.003), RBV (8,424 ± 5,405) (p = 0.028), and MTT (262 ± 90.6 s) (p = 0.005). In contrast, the parameter of regional blood flow (32.1 ± 10.9) was significantly lower (p = 0.038). The peak signal intensity (25.8 ± 8.2 %) was lower, but this was not significant (p = 0.083).
Conclusions:
CEUS provides a reliable non-invasive imaging modality for the assessment of lower limb muscle perfusion pressures. This may be of clinical use in the assessment of a developing compartment syndrome. Further clinical studies are required to further define its accuracy and reproducibility.
More Related Videos
08:22Combined Intravital Microscopy and Contrast-enhanced Ultrasonography of the Mouse Hindlimb to Study Insulin-induced Vasodilation and Muscle Perfusion
Published on: March 20, 2017
12:15Tissue Preparation Techniques for Contrast-Enhanced Micro Computed Tomography Imaging of Large Mammalian Cardiac Models with Chronic Disease
Published on: February 8, 2022
Related Concept Videos
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for...
Ultrasonography
During an ultrasonography procedure, a handheld device called...
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion,...