Related Experiment Video
Updated: Mar 26, 2026

Modified Posterior Vertebral Column Resection for Patients with Thoracolumbar Kyphotic Deformity
Published on: September 16, 2022
Secondary Collapse of an Expandable Cage After Vertebral Corpectomy
Martin Sattler1, Thomas Goesling, Marc Busche
1Trauma Department, Hannover Medical School, Carl-Neuberg-Strasse 1, 30625, Hannover, Germany. sattler.martin@mh-hannover.de.
Abstract:
Expandable vertebral body replacement systems have been increasingly used for anterior stabilization of spine. We report a secondary collapse of an expandable vertebral body replacement system. This specific complication has not been reported in the literature so far. The most obvious reason for failure was insufficient tightening of a locking screw. This paper emphasizes the importance of correct technical application.