Optic nerve sheath measurement and raised intracranial pressure in paediatric traumatic brain injury

S Agrawal1,2, J Brierley3

  • 1Paediatric and Neonatal Intensive Care Unit, Institute of Child Health and Great Ormond Street Hospital for Children, London, WC1N 3JH, UK. shragrawal@gmail.com.

Insights

Optic nerve sheath diameter (ONSD) ultrasound can help identify children with traumatic brain injury (TBI) who do not have elevated intracranial pressure (ICP). This non-invasive method may avoid unnecessary invasive ICP monitoring in pediatric intensive care units.

Area of Science:

  • Pediatric Neurocritical Care
  • Neuroimaging
  • Trauma Surgery

Background:

  • Management of severe traumatic brain injury (TBI) in children often requires invasive intracranial pressure (ICP) monitoring.
  • However, not all TBI patients develop elevated ICP, indicating a need for non-invasive screening methods.
  • Identifying children with normal ICP can help avoid unnecessary invasive procedures.

Purpose of the Study:

  • To evaluate the utility of optic nerve sheath diameter (ONSD) measurement as a non-invasive screening tool for elevated ICP in children with severe TBI.
  • To determine if ONSD can reliably differentiate between children with and without elevated ICP.

Main Methods:

  • Ultrasound measurement of ONSD using a 10-MHz probe in 11 children (age 2-15 years) with severe TBI requiring ICP monitoring.
  • Simultaneous invasive ICP monitoring was performed; ICP >15 mmHg was considered abnormal.
  • An ONSD >4.5 mm in children over 1 year was considered abnormal.

Main Results:

  • All children with clinically significant elevated ICP demonstrated abnormal ONSD measurements.
  • Children with normal ICP had normal ONSD findings.
  • The study included a small cohort of 11 pediatric patients.

Conclusions:

  • Optic nerve sheath diameter (ONSD) measurement shows promise as a non-invasive method to identify children with TBI and normal ICP.
  • This approach may help clinicians avoid the insertion of unnecessary invasive ICP monitors.
  • Further research with larger cohorts is warranted to confirm these findings.
Abstract