Related Experiment Video
Updated: Mar 26, 2026

Ultrasonographic Assessment During Cardiopulmonary Resuscitation
Published on: October 24, 2020
Tracheal Tube Position Shift During Infant Resuscitation by Chest Compression: A Simulation Comparison by Fixation
Takeshi Ueno1, Nobuyasu Komasawa1, Nozomi Majima1
1Department of Anesthesiology, Osaka Medical College, Osaka, Japan.
Background:
Tracheal tube placement during infant resuscitation is essential for definite airway protection. Accidental extubation due to tracheal tube displacement is a rare event, but it results in severe complications, especially in infants.
Objective:
The present study evaluated how infant tracheal tube displacement is affected by tape vs. tube holder fixation using a manikin.
Methods:
A tracheal tube with internal diameter of 3.5 mm was placed 10 cm from the gum ridge in an advanced life support (ALS) Baby(®) simulator (Laerdal, Stavanger, Norway). In the first trial, cuff pressure was set at 15, 20, and 25 cmH2O and trials were performed at each setting with no fixation, Durapore(®) (3M, St Paul, MN) tape fixation, Multipore(®) (3M) tape fixation, and Thomas(®) Tube Holder (Laerdal) fixation. After 5 min of chest compression, the tracheal tube shift was measured. In the second trial, we compared the tube shift by chest compression with or without cuff in the same way.
Results:
Relative to no fixation, tracheal tube shift was significantly less in the Durapore, Multipore, and tube holder groups (p < 0.05) at all cuff settings. Of the three fixation methods, the tube holder showed significantly less shift (p < 0.05) relative to tape, regardless of the initial cuff pressure. The positional shift after chest compressions was significantly larger in the trials with cuff than in those without cuff in Durapore or Multipore fixation (p < 0.05), but did not in tube holder fixation.
Conclusions:
There is less tracheal tube displacement with tube holder fixation than with tape during continuous infant chest compression simulation. The tube cuff can contribute to the positional shift of the tube during infant chest compression.
Related Concept Videos
Cardiopulmonary Resuscitation II: ACLS Airway Management
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation: