Related Experiment Video
Updated: Mar 26, 2026

Surgical Technique for the Implantation of a Biomimetic Artificial Intervertebral Disc in a Goat Animal Model
Published on: October 10, 2025
A Biomechanical Analysis of an Artificial Disc With a Shock-absorbing Core Property by Using Whole-cervical Spine
June Ho Lee1, Won Man Park2, Yoon Hyuk Kim2
1Department of Neurosurgery, Kyung Hee University Medical Center, Seoul.
Study Design:
A biomechanical comparison among the intact C2 to C7 segments, the C5 to C6 segments implanted with fusion cage, and three different artificial disc replacements (ADRs) by finite element (FE) model creation reflecting the entire cervical spine below C2.
Objective:
The aim of this study was to analyze the biomechanical changes in subaxial cervical spine after ADR and to verify the efficacy of a new mobile core artificial disc Baguera C that is designed to absorb shock.
Summary Of Background Data:
Scarce references could be found and compared regarding the cervical ADR devices' biomechanical differences that are consequently related to their different clinical results.
Methods:
One fusion device (CJ cage system, WINNOVA) and three different cervical artificial discs (Prodisc-C Nova (DePuy Synthes), Discocerv (Scient'x/Alphatec), Baguera C (Spineart)) were inserted at C5-6 disc space inside the FE model and analyzed. Hybrid loading conditions, under bending moments of 1 Nm along flexion, extension, lateral bending, and axial rotation with a compressive force of 50 N along the follower loading direction, were used in this study. Biomechanical behaviors such as segmental mobility, facet joint forces, and possible wear debris phenomenon inside the core were investigated.
Results:
The segmental motions as well as facet joint forces were exaggerated after ADR regardless of type of the devices. The Baguera C mimicked the intact cervical spine regarding the location of the center of rotation only during the flexion moment. It also showed a relatively wider distribution of the contact area and significantly lower contact pressure distribution on the core than the other two devices. A "lift off" phenomenon was noted for other two devices according to the specific loading condition.
Conclusion:
The mobile core artificial disc Baguera C can be considered biomechanically superior to other devices by demonstrating no "lift off" phenomenon, and significantly lower contact pressure distribution on core.
Level Of Evidence:
N/A.
Related Concept Videos
Design of Transmission Shafts - Stress Analysis
General Case of Eccentric Axial Loading
Consider a member subjected to equal and opposite forces that are applied along a line that does not coincide with the member's neutral axis. In unsymmetrical...
Normal Strain under Axial Loading
Eccentric Axial Loading in a Plane of Symmetry