Multi-Scale Modification of Metallic Implants With Pore Gradients, Polyelectrolytes and Their Indirect Monitoring In vivo
Published on: July 1, 2013
Surface characterization and in vivo evaluation of laser sintered and machined implants followed by
Michelle Bowers1, Daniel Yoo, Charles Marin
1345 E 24th Street, 10010, Biomaterials and Biomimetics, New York University, New York, NY, USA, rh1694@@nyu.edu.
Background:
This study aimed to compare the histomorphometric and histological bone response to laser-sintered implants followed by resorbable-blasting media (RBM) process relative to standard machined/RBM surface treated implants.
Material And Methods:
Six male sheep (n=6) received 2 Ti-6Al-4V implants (1 per surface) in each side of the mandible for 6 weeks in vivo. The histomorphometric parameters bone-implant contact (BIC) and bone area fraction occupancy (BAFO) were evaluated.
Results:
Optical interferometry revealed higher Sa and Sq values for the laser-sintered/RBM surface in relation to standard/RBM implants. No significant differences in BIC were observed between the two groups (p>0.2), but significantly higher BAFO was observed for standard/RBM implants (p<0.01).
Conclusions:
The present study demonstrated that both surfaces were biocompatible and osseoconductive, and the combination of laser sintering and RBM has no advantage over the standard machined implants with subsequent RBM.