Measurement of Renal Depth in Dynamic Renal Scintigraphy Using Ultralow-Dose CT
Yusuke Inoue1, Hiroyasu Itoh, Hirokazu Tagami
1From the *Department of Diagnostic Radiology, Kitasato University School of Medicine; and †Department of Radiology, Kitasato University Hospital, Sagamihara, Japan.
Purpose:
Renal depths predicted using predefined formulas are commonly used for camera-based evaluation of renal function. We investigated the feasibility and utility of renal depth measurement using ultralow-dose CT images acquired in conjunction with dynamic renal scintigraphy.
Methods:
Dynamic renal scintigraphy with Tc-MAG3 was performed in 117 patients (225 kidneys) using a SPECT/CT scanner, and ultralow-dose CT (estimated effective dose of 0.17 mSv) was performed during free breathing immediately before tracer injection. The clarity of the renal contour on the CT images was evaluated visually. The renal depths were measured by 2 methods and compared with depths predicted by 2 previously reported methods. The accuracy of camera-based clearance using predicted and measured depths was evaluated using a single-sample method as a standard.
Results:
The clarity of the renal contour was poor in 18 of 225 kidneys, and 12 of 117 patients were considered ineligible for depth measurement. The measurement for eligible patients showed excellent intraobserver and interobserver repeatabilities. Although mean depths were similar among the 2 CT measurement methods and 2 prediction methods, absolute differences of more than 1 cm were observed in approximately 20% of kidneys between CT measurement and prediction. CT measurement of renal depth failed to improve the accuracy of camera-based clearance evaluation.
Conclusion:
Ultralow-dose CT allowed measurement of renal depth in most patients. Substantial differences in renal depth between prediction and CT measurement indicated potential usefulness of CT measurement, although no actual improvement in the accuracy of clearance estimation was demonstrated in this study.
More Related Videos
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
03:19Author Spotlight: Developing a Bedside Protocol for Kidney and Genitourinary Ultrasonography
Published on: June 21, 2024
Related Concept Videos
Imaging Studies III: Computed Tomography
Imaging Studies II: Ultrasonography
Drug Dosing in Renal Diseases: Measurement of Glomerular Filtration Rate
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System V: CT
Imaging Studies VII: Vascular Imaging