Accuracy of QT interval measurement on electrocardiographs displayed on electronic 'smart' devices
Joe-Anthony Rotella1,2, David McD Taylor2, Anselm Wong1,2
1Victorian Poisons Information Centre, Austin Health, Melbourne, Victoria, Australia.
Objective:
The objective of the study was to compare QT intervals measured on the original bedside electrocardiographs (ECG), facsimile, iPhone, iPad and 17 inch computer monitor.
Methods:
This was a prospective, observational study conducted within a tertiary referral metropolitan ED. Thirteen doctors measured the QT intervals of 15 non-identifiable, routinely recorded ECGs using randomly allocated modalities over five sessions. The QT intervals of lead I, II, V2, V4, V6 and one other were measured. Variations between modalities and between doctors were examined (χ(2) and Kruskal-Wallis tests).
Results:
QT intervals measured on the iPad were significantly lower than the other modalities (P = 0.02). For each lead, there were no differences in QT interval measurement between the modalities (P > 0.05). However, there were significant differences between the QT interval measurements of the participating doctors (range of median QT intervals 480-530 msec, P < 0.001). There were also significant differences between the QT intervals measurement of the leads examined (range of median QT intervals 480-520 msec, P < 0.001). Lead V2 was the most consistent lead (smallest interquartile range), and lead I was the least (QT interval consistently under-estimated).
Conclusion:
The QT interval can be reliably measured using facsimile, iPhone and computer, but not the iPad. However, it varies between doctors. Lead V2 is recommended for QT interval measurement if only a single lead is to be used for clinical purposes.
Related Concept Videos
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...
ECG Interpretation of Rhythms
Components of the Electrocardiogram
The primary components of a normal ECG waveform in Normal sinus rhythm(NSR) include the P wave, PR interval, QRS complex, ST segment, T wave, and occasionally a U wave.
ECG waveforms are divided by vertical and horizontal lines at standard intervals.
The horizontal axis measures time and rate, and the vertical axis measures amplitude or voltage....
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin...
Holter Monitor: 24-Hour Monitoring
Dysrhythmias V: Evaluating Dysrhythmias
Correlation between ECG and Cardiac Cycle
A cardiac action potential originates in the SA node and spreads throughout the atria and the AV node in approximately 0.03 seconds. This results in the P wave in an ECG and triggers atrial contraction. The action potential is then briefly slowed at the AV node, allowing the atria to contract and fill the ventricles with blood before...