Multi-Scale Modification of Metallic Implants With Pore Gradients, Polyelectrolytes and Their Indirect Monitoring In vivo
Published on: July 1, 2013
In vivo comparative study of tissue reaction to bare and antimicrobial polymer coated transcutaneous implants
Tilman Calliess1, Ivonne Bartsch1, Maike Haupt1
1Laboratory for Biomechanics and Biomaterials, Hannover Medical School, Anna-von-Borries-Straße 1-7, 30625 Hannover, Germany; CrossBIT, Center for Biocompatibility and Implant-Immunology, Feodor-Lynen-Straße 31, 30625 Hannover, Germany.
Abstract:
We coated transcutaneous implants made of titanium alloy Ti6Al4V with copolymer dimethyl (2-methacryloyloxy-ethyl) phosphonate and 4-vinylpyridine and investigated the tissue reaction with respect to its biocompatible and antimicrobial properties in vivo. We distinguished between clinically observable superficial inflammations and histologically detectable deep infections. The vinylpyridine moieties were transferred into cationic pyridinium groups by reaction with hexyl bromide. Thus polymers with both antimicrobial capacity and good biocompatibility were obtained. In a short-term study, we implanted specially designed bare or coated implants in hairless but immunocompetent mice and analyzed the tissue reaction histologically. No difference was found between bare and coated implants in the initial healing phase of up to 14 days; however, after 21 days the scar tissue formation was higher in the bare implant group. The degree of epithelial downgrowth was comparable in both groups at any time point. In a long-term study of up to 168 days, we analyzed resistance to infection. In the bare implant group, 7 of the 12 implantation sites became infected deep whereas in the coated implant group only two deep infections were observed. The other implantation sites showed only superficial signs of inflammation. These results generally accord with previous in-vitro studies.