Coping with real world data: Artifact reduction and denoising for motion-compensated cardiac C-arm CT
Oliver Taubmann1, Andreas Maier1, Joachim Hornegger1
1Pattern Recognition Lab, Computer Science Department, Friedrich-Alexander-University Erlangen-Nuremberg, 91058 Erlangen, Germany and Erlangen Graduate School in Advanced Optical Technologies (SAOT), 91052 Erlangen, Germany.
Purpose:
Detailed analysis of cardiac motion would be helpful for supporting clinical workflow in the interventional suite. With an angiographic C-arm system, multiple heart phases can be reconstructed using electrocardiogram gating. However, the resulting angular undersampling is highly detrimental to the quality of the reconstructed images, especially in nonideal intraprocedural imaging conditions. Motion-compensated reconstruction has previously been shown to alleviate this problem, but it heavily relies on a preliminary reconstruction suitable for motion estimation. In this work, the authors propose a processing pipeline tailored to augment these initial images for the purpose of motion estimation and assess how it affects the final images after motion compensation.
Methods:
The following combination of simple, direct methods inspired by the core ideas of existing approaches proved beneficial: (a) Streak reduction by masking high-intensity components in projection domain after filtering. (b) Streak reduction by subtraction of estimated artifact volumes in reconstruction domain. (c) Denoising in spatial domain using a joint bilateral filter guided by an uncompensated reconstruction. (d) Denoising in temporal domain using an adaptive Gaussian smoothing based on a novel motion detection scheme.
Results:
Experiments on a numerical heart phantom yield a reduction of the relative root-mean-square error from 89.9% to 3.6% and an increase of correlation with the ground truth from 95.763% to 99.995% for the motion-compensated reconstruction when the authors' processing is applied to the initial images. In three clinical patient data sets, the signal-to-noise ratio measured in an ideally homogeneous region is increased by 37.7% on average. Overall visual appearance is improved notably and some anatomical features are more readily discernible.
Conclusions:
The authors' findings suggest that the proposed sequence of steps provides a clear advantage over an arbitrary sequence of individual image enhancement methods and is fit to overcome the issue of lacking image quality in motion-compensated C-arm imaging of the heart. As for future work, the obtained results pave the way for investigating how accurately cardiac functional motion parameters can be determined with this modality.
More Related Videos
11:09High-Resolution Cardiac Positron Emission Tomography/Computed Tomography for Small Animals
Published on: December 16, 2022
06:53Management of Respiratory Motion Artefacts in 18F-fluorodeoxyglucose Positron Emission Tomography using an Amplitude-Based Optimal Respiratory Gating Algorithm
Published on: July 23, 2020
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...