A Low Cost Setup for Behavioral Audiometry in Rodents
Published on: October 16, 2012
Behavioral Pure-Tone Threshold Shifts Caused by Tympanic Membrane Electrodes
Spencer B Smith1, Jeffery Lichtenhan, Barbara Cone
11Department of Speech, Language, and Hearing Sciences, University of Arizona, Tucson, Arizona, USA; and 2Department of Otolaryngology, Washington University in St. Louis School of Medicine, St. Louis, Missouri, USA.
Objective:
To determine whether tympanic membrane (TM) electrodes induce behavioral pure-tone threshold shifts.
Design:
Pure-tone thresholds (250 to 8000 Hz) were measured twice in test (n = 18) and control (n = 10) groups. TM electrodes were placed between first and second threshold measurements in the test group, whereas the control group did not receive electrodes. Pure-tone threshold shifts were compared between groups. The effect of TM electrode contact location on threshold shifts was evaluated in the test group.
Results:
TM electrodes significantly increased average low-frequency thresholds, 7.5 dB at 250 Hz and 4.2 dB at 500 Hz, and shifts were as large as 25 dB in individual ears. Also, threshold shifts did not appear to vary at any frequency with TM electrode contact location.
Conclusions:
Low-frequency threshold shifts occur when using TM electrodes and insert earphones. These findings are relevant to interpreting electrocochleographic responses to low-frequency stimuli.
More Related Videos
14:05Behavioral Assessment of Hearing in 2 to 4 Year-old Children: A Two-interval, Observer-based Procedure Using Conditioned Play-based Responses
Published on: January 23, 2017
11:39Assessment of Audio-Tactile Sensory Substitution Training in Participants with Profound Deafness Using the Event-Related Potential Technique
Published on: September 7, 2022
Related Concept Videos
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
NMR Spectroscopy: Chemical Shift Overview
For instance, the proton...