[Multiple inappropriate defibrillator shocks due to insulation failure of a Biotronik Linox defibrillator lead with
Hamdi Elfarra1, Rainer Moosdorf1, Leszek Rybinski1
1Zentrum für Herz- und thorakale Gefäßchirurgie, Philipps-Universität Marburg, Marburg, Deutschland.
Abstract:
In this article the case of a patient who received a total of 35 inappropriate defibrillator shocks due to insulation failure with externalized conductor of a Biotronik Linox® lead is described. The implanted defibrillator was immediately inactivated and the failed lead was extracted using a laser sheath system.