Do magnetic growing rods have lower complication rates compared with conventional growing rods?
Kar H Teoh1, Daniel M G Winson1, Stuart H James1
1Welsh Centre for Spinal Surgery & Trauma, University Hospital of Wales, Heath Park, Cardiff CF10 2FF, United Kingdom.
Background Context:
The main advantage cited for the use of the magnetic controlled growing rod (MCGR) system over the conventional growing rod (CGR) in early-onset scoliosis is avoiding repeated invasive surgical procedures for lengthening, thus reducing, complications.
Purpose:
The study aimed to evaluate the complications of the MCGR system against the CGR system in our center.
Study Design/Setting:
This is a retrospective case control series.
Patient Sample:
The sample includes patients with early-onset scoliosis treated with MCGR or CGR.
Outcome Measures:
Complications and unplanned return to theater were the outcome measures.
Results:
Of the 37 patients (MCGR, N=10; CGR, N=27) in our cohort, 28 patients (76%) had at least one complication. Taking into account the follow-up period, MCGR had a higher complication rate than CGR group (0.32 complication per patient per year vs. 0.15 complication per patient per year). The use of MCGR was associated with a lower risk of deep infection (odds ratio [OR]: 0.22; p=.22) and superficial infection (OR: 0.07, p=.017) but increased risk of metalwork problems (OR: 4.67; p=.045) and unplanned return to theater (OR: 2.92; p=.05) compared with CGR.
Conclusions:
Although MCGR has a lower rate of both deep and superficial infections when compared with CGR, it does not completely avoid repeated invasive surgical procedures as previously suggested. It does have a significant increased risk of metalwork problems and unplanned return to theater.
Related Concept Videos
Thermal Strain
Thermal expansion and Thermal stress: Problem Solving
To solve the problem, first, identify the known and unknown quantities. The initial length (L) of the bridge is 1275 m, the coefficient of linear expansion (α) for steel is 12 x 10-6/°C, and the change in temperature (ΔT) is 55...
Motional Emf
Normal Strain under Axial Loading
Magnetic Field Due to Two Straight Wires
Magnetic Susceptibility and Permeability
When diamagnetic materials are placed under an external magnetic field, the moments opposite to the field are induced. Hence, the susceptibility for diamagnets has a minimal negative value of 10-5–10-6. Since...