Comparing smartphone camera adapters in imaging post-operative cataract patients
Thuss Sanguansak1, Katharine Morley2, Michael Morley3
11 Faculty of Medicine, Khon Kaen University, Thailand.
Journal of Telemedicine and Telecare
|February 7, 2016
Summary
Smartphone anterior segment imaging using macro lens with LED illumination (ML-LED), no adapter (NA), and slit lamp (SL) adapters provide high image quality and clinical confidence for post-operative cataract patients.
Area of Science:
- Ophthalmology
- Medical Imaging
- Mobile Health Technology
Background:
- Smartphone-based anterior segment imaging is emerging as a tool for ophthalmological assessments.
- Evaluating image quality and clinical utility of different smartphone adapters is crucial for real-world application.
Purpose of the Study:
- To compare image quality and clinical confidence using four smartphone anterior segment imaging adapters: macro lens (ML), ML with LED illumination (ML-LED), no adapter (NA), and slit lamp (SL).
Main Methods:
- Anterior segment images were captured from 190 post-cataract surgery eyes using an iPhone 6 with ML, ML-LED, NA, and SL adapters.
- Ophthalmologists subjectively rated images for quality and clinical management confidence.
Main Results:
- No adapter (NA), ML-LED, and SL adapters achieved high image quality acceptability (100%, 93.7%, 86.3%).
- NA, SL, and ML-LED demonstrated high clinical confidence (100%, 98%, 93.2%).
- The macro lens (ML) adapter showed the lowest acceptability for both image quality (61.1%) and clinical confidence (37.4%).
Conclusions:
- ML-LED, NA, and SL adapters are suitable for visualizing anterior segment structures, achieving 86-100% acceptability.
- These smartphone imaging methods, combined with clinical data, offer 93-100% acceptable clinical confidence for managing post-operative cataract patients.