Related Experiment Video
Updated: Mar 26, 2026

Author Spotlight: Developing a Bedside Protocol for Kidney and Genitourinary Ultrasonography
Published on: June 21, 2024
Intrarenal pressure and irrigation flow with commonly used ureteric access sheaths and instruments
Anna Wright1, Kevin Williams1, Bhaskar Somani2
1University Hospital Southampton NHS Trust, Department of Urology, Southampton, United Kingdom.
Introduction:
Flexible ureterorenoscopy is becoming a first-line treatment for many intrarenal stones. Ureteric access sheaths are commonly used to aid access, stone removal and reduce intrarenal pressure. We evaluated the effects of two commonly used access sheaths on irrigation flow and intrarenal pressure during flexible ureterorenoscopy. We measured the effect of scope instrumentation on flow and pressure.
Material And Methods:
We utilized a 10/12F and 12/14F, 35 cm Re-Trace™ access sheath with a FlexX2 scope in a cadaveric porcine kidney. We evaluated the effect of four Nitinol baskets (1.3F, 1.5F, 1.9F, 2.2F), three different 200 µm laser fibres and a hand-held pump. Measurements of irrigation flow and intrarenal pressure were recorded and compared between the different sized access sheaths.
Results:
Flow rates varied widely between access sheaths. Without instrumentation, mean flow was 17 mls/min (10/12F access sheath), versus 33 mls/min (12/14F sheath) (p <0.0001). Increasing basket size produced a gradual reduction in flow and pressure in both access sheaths. Reassuringly, pressures were low overall (<40 cm H2O). Pressures were significantly reduced when using the larger 12/14F sheath, with and without all instrumentations (p <0.0001). Hand-held pump devices have a marked effect on flow and pressure in both sheaths; with pressures rising up to 121 cm H2O with a 10/12F sheath, versus 29 cm H2O (12/14F) (p <0.0001).
Conclusions:
A 12/14F access sheath offered significantly improved irrigation whilst maintaining significantly lower intrarenal pressure, when compared to a 10/12F access sheath in a cadaveric porcine model. Scope instrumentation affects irrigation flow and pressure in both sized sheaths. Furthermore, there should be caution with hand-held pump devices, especially with smaller sized sheaths, as intrarenal pressure can be very high.
More Related Videos
06:39Technical Modification of the Terminal Ureter During Total Transperitoneal Laparoscopic Nephroureterectomy for Upper Urinary Tract Urothelial Carcinoma
Published on: November 22, 2019
09:17In Vivo Luminal Measurement of Distension-Evoked Urothelial ATP Release in Rodents
Published on: September 7, 2022
Related Concept Videos
Imaging Studies V: Intravenous Urography and Retrograde Pyelography
Imaging Studies II: Ultrasonography
Imaging Studies VII: Vascular Imaging
Urinary Tract Calculi V: Nursing Management
Urinary Tract Calculi VI: Surgical Management
Acute Kidney Injury IV: Diagnostic Studies and Prevention