Accuracy of Measurements With the iCare HOME Rebound Tonometer
Julia Termühlen1, Natasa Mihailovic, Maged Alnawaiseh
1*Department of Ophthalmology, University of Münster, Münster †Center of Ophthalmology, University of Cologne, Cologne ‡Department of Ophthalmology, University of Würzburg, Würzburg, Germany.
Purpose:
To evaluate the accuracy of intraocular pressure (IOP) measurements obtained with the newly available iCare HOME (RTHOME) rebound tonometer compared with the iCare ONE (RTONE) tonometer and Goldmann applanation tonometry (GAT), and possible correlation with central corneal thickness (CCT).
Materials And Methods:
IOP measurements were obtained from 154 patients by an ophthalmologist (doc) using each of the above-mentioned tonometers. In addition, patients (pat) measured their own IOP with the RTHOME and RTONE. The means and SD of results obtained with the different tonometers were compared. Agreement between the tonometers was calculated using the Bland-Altman method.
Results:
Mean IOPs for the right eyes only were 15.9±6.4 mm Hg (RTONEdoc), 15.8±6.4 mm Hg (RTONEpat), 15.0±5.9 mm Hg (RTHOMEdoc), 14.9±6.3 mm Hg (RTHOMEpat), and 15.8±4.4 mm Hg (GAT). Bland-Altman analysis revealed mean differences (bias) between RTONEdoc and RTHOMEdoc, between RTHOMEdoc and RTHOMEpat, and between RTHOMEdoc and GAT of 0.8, 0.1, and -0.8 mm Hg, respectively, with 95% limits of agreement of -3.5 to 5.2, -4.9 to 5.1, and -7.2 to 5.6 mm Hg, respectively. Linear regression of the comparisons revealed a proportional error over the range of pressures examined in the case of RTHOMEdoc versus GAT (slope=0.32, P<0.001). Considering the data from all eyes, the difference between RTHOMEdoc and GAT correlated significantly with the CCT (P=0.01).
Conclusion:
RTHOME readings correlate well with the GAT results although some limitations such as dependency of readings on CCT and increasing differences at lower and higher IOP levels need to be taken into account.
More Related Videos
05:14Comparison of Agreement and Accuracy using Binocular Wavefront Optometer with Autorefractor and Phoropter
Published on: September 16, 2025
08:40Isokinetic Robotic Device to Improve Test-Retest and Inter-Rater Reliability for Stretch Reflex Measurements in Stroke Patients with Spasticity
Published on: June 12, 2019
Related Concept Videos
Measurement of Blood Pressure
Uncertainty in Measurement: Reading Instruments
Assessment of blood pressure in brachial artery(two-step method)
Errors occurring during blood pressure monitoring
Several factors...
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.