Intraoperative Real-time Cochlear Response Telemetry Predicts Hearing Preservation in Cochlear Implantation
Luke Campbell1, Arielle Kaicer, David Sly
1*Department of Otolaryngology, University of Melbourne †Royal Victorian Eye and Ear Hospital, East Melbounre, VIC, Australia.
Aim:
To monitor cochlear function during cochlear implantation and determine correlations with postoperative acoustic hearing.
Background:
Cochlear response telemetry measures cochlear function directly from cochlear implant electrodes. We have adapted this system to provide real-time cochlear response telemetry (RT-CRT) monitoring of a patient's acoustic hearing as the cochlear implant electrode array is inserted.
Methods:
Eighteen subjects (1 child and 17 adults) with sloping high frequency hearing loss were implanted with Cochlear Ltd slim straight arrays (CI422/CI522). Tone bursts (500 Hz, 100-110 dB) were presented at 14 Hz continuously during the array insertion. RT-CRT amplitudes were correlated with surgical manoeuvres recorded on the video from the operating microscope and with postoperative pure tone audiograms.
Results:
Despite an excellent overall rate of complete or partial hearing preservation (79%), RT-CRT identified that in 47% of these implantations there was transient or permanent reduction in the amplitude of the cochlear microphonic (CM). Patients with a preserved CM at the end of insertion had on average 15 dB better low-frequency hearing preservation. The CM amplitude was most vulnerable during the last few millimeters of insertion or when inadvertent movement of the array occurred after full insertion. Physical contact/elevation of the basilar membrane is hypothesized as a likely mechanism of hearing loss rather than overt physical trauma.
Conclusion:
RT-CRT can be used to predict early postoperative hearing loss and to potentially refine surgical technique. In the future, feedback of RT-CRT may prove to be a valuable tool for maximizing preservation of residual hearing or providing feedback on electrode contact with the basilar membrane.