Background:
Current investigations of stomach function are based on small test meals that do not reliably induce symptoms and analysis techniques that rarely detect clinically relevant dysfunction. This study introduces the large 'Nottingham Test Meal' (NTM) for assessment of gastric motor and sensory function by non-invasive imaging.
Methods:
NTM comprises 400 mL liquid nutrient (0.75 kcal/mL) and 12 solid agar-beads (0 kcal) with known breaking strength. Gastric fullness and dyspeptic sensations were documented by 100 mm visual analogue scale (VAS). Gastric emptying (GE) were measured in 24 healthy volunteers (HVs) by gastric scintigraphy (GS) and magnetic resonance imaging (MRI). The contribution of secretion to gastric volume was assessed. Parameters that describe GE were calculated from validated models. Inter-observer agreement and reproducibility were assessed.
Key Results:
NTM produced moderate fullness (VAS ≥30) but no more than mild dyspeptic symptoms (VAS <30) in 24 HVs. Stable binding of meal components to labels in gastric conditions was confirmed. Distinct early and late-phase GE were detected by both modalities. Liquid GE half-time was median 49 (95% CI: 36-62) min and 68 (57-71) min for GS and MRI, respectively. Differences between GS and MRI measurements were explained by the contribution of gastric secretion. Breaking strength for agar-beads was 0.8 N/m(2) such that median 25 (8-50) % intact agar-beads and 65 (47-74) % solid material remained at 120 min on MRI and GS, respectively. Good reproducibility for liquid GE parameters was present and GE was not altered by agar-beads.
Conclusions & Inferences:
The NTM provided an objective assessment of gastric motor and sensory function. The results were reproducible and liquid emptying was not affected by non-nutrient agar-beads. The method is potentially suitable for clinical practice.
More Related Videos
09:29Fabrication and Implantation of Miniature Dual-element Strain Gages for Measuring In Vivo Gastrointestinal Contractions in Rodents.
Published on: September 18, 2014
12:24Assessment of Gastric Emptying in Non-obese Diabetic Mice Using a [13C]-octanoic Acid Breath Test
Published on: March 23, 2013
Related Concept Videos
Imaging Studies III: Gastrointestinal Motility Studies and Virtual Colonoscopy
Radionuclide Testing
Radionuclide testing is a sophisticated medical technique for assessing gastrointestinal motility. It focuses on gastric emptying and colonic transit time. Radioactive markers track the movement of food through the digestive system, providing insights into gastrointestinal disorders.
In gastric emptying studies, a meal's liquid and...
Gastric Emptying
Gastric Motility
Peristaltic Waves and Chyme Formation
Upon food entry, the stomach initiates...
Assessment of the Gastrointestinal System I: Subjective Data
Health History
The initial step in assessing the GI system is obtaining a comprehensive health history. This includes inquiring about the patient's history or presence of problems...
Assessment of the Gastrointestinal System II: Health Perception Pattern
Health Perception Patterns
Health perception patterns offer valuable insights into a patient's lifestyle habits and how they may impact their GI health. These patterns include:
Gastrointestinal Motility Disorders