StatNet Electroencephalogram: A Fast and Reliable Option to Diagnose Nonconvulsive Status Epilepticus in Emergency
Lady Diana Ladino1, Alexandra Voll1, Dianne Dash1
11Department of Medicine,Division of Neurology,University of Saskatchewan,Saskatoon,Canada.
Background:
The StatNet electrode set is a system that can be applied by a non-electroencephalogram (EEG) technologist after minimal training. The primary objectives of this study are to assess the quality and reliability of the StatNet recordings in comparison to the conventional EEG.
Methods:
Over 10 months, 19 patients with suspected nonconvulsive status epilepticus were included from university hospital emergency settings. Each patient received a StatNet EEG by a trained epilepsy fellow and a conventional EEG by registered technologists. We compared the studies in a blinded fashion, for the timeframe from EEG order to the setup time, start of acquisition, amount of artifact, and detection of abnormalities. The nonparametric Mann-Whitney two-sample t test was used for comparisons. The kappa score was used to assess reliability.
Results:
Mean age of patients was 61±16.3 (25-93) years. The inter-observer agreement for detection of abnormal findings was 0.83 for StatNet and 0.75 for conventional EEG. Nonconvulsive status epilepticus was detected in 10% (2/19) in both studies. The delay from the time of EEG requisition to acquisition was shorter in the StatNet (22.4±2.5 minutes) than the conventional EEG (217.7±44.6 minutes; p<0.0001). The setup time was also shorter in the StatNet (9.9±0.8 minutes) compared with the conventional EEG (17.8±0.8 minutes; p<0.0001). There was no difference in the percentage of artifact duration between the two studies (p=0.89).
Conclusion:
This study demonstrates that StatNet EEG is a practical and reliable tool in the emergency setting, which reduces the delay of testing compared with conventional EEG, without significant compromise of study quality.