Setup error analysis in helical tomotherapy based image-guided radiation therapy treatments.
Bhagyalakshmi Akkavil Thondykandy1, Jamema V Swamidas2, Jayprakash Agarwal3
1Department of Medical Physics, Tata Memorial Hospital, Mumbai, Maharashtra, India.
Journal of Medical Physics
|February 12, 2016
Summary
Setup margins for helical tomotherapy were evaluated. Clinical margins are adequate for brain, head and neck, and lung sites, but insufficient for craniospinal irradiation and pelvis treatments.
Area of Science:
- Radiation Oncology
- Medical Physics
Background:
- Accurate patient positioning is crucial in helical tomotherapy for effective cancer treatment.
- Setup margins are essential to account for systematic and random errors during radiation delivery.
Purpose of the Study:
- To investigate the adequacy of clinical setup margins for various anatomical sites treated with helical tomotherapy.
- To compare calculated margins with clinically used margins to identify potential discrepancies.
Main Methods:
- Analysis of 2686 megavoltage computed tomography images from 102 patients across brain, head and neck, pelvis, lung, and craniospinal irradiation (CSI) sites.
- Calculation of systematic and random errors in translational and rotational directions using the van Herk formula.
- Comparison of calculated setup margins with established clinical margins for each site.
Main Results:
- Clinical margins were adequate for brain, head and neck (H and N), and lung sites.
- Longitudinal margins for brain patients were slightly out of tolerance (0.7 mm).
- Pelvic and CSI sites showed discrepancies, with calculated margins exceeding clinical allowances in certain directions, indicating potential undercoverage.
Conclusions:
- Current clinical setup margins are sufficient for H and N, lung, and brain treatments.
- Helical tomotherapy treatments for pelvis and CSI require re-evaluation of setup margins to ensure adequate coverage and patient safety.
- Site-specific margin adjustments may be necessary to optimize treatment accuracy.