Human Factors Approach to Comparative Usability of Hospital Manual Defibrillators
Richard Fidler1, Meshell Johnson2
1San Francisco Veterans Affairs Medical Center, University of California, San Francisco, CA 94121, United States; School of Nursing, Departments of Anesthesia and Perioperative Medicine, University of California, San Francisco, CA 94121, United States.
Introduction:
Equipment-related issues have recently been cited as a significant contributor to the suboptimal outcomes of resuscitation management. A systematic evaluation of the human-device interface was undertaken to evaluate the intuitive nature of three different defibrillators. Devices tested were the Physio-Control LifePak 15, the Zoll R Series Plus, and the Philips MRx.
Methods:
A convenience sample of 73 multidisciplinary health care providers from 5 different hospitals participated in this study. All subjects' performances were evaluated without any training on the devices being studied to assess the intuitiveness of the user interface to perform the functions of delivering an Automated External Defibrillator (AED) shock, a manual defibrillation, pacing to achieve 100% capture, and synchronized cardioversion on a rhythm simulator.
Results:
Times to deliver an AED shock were fastest with the Zoll, whereas the Philips had the fastest times to deliver a manual defibrillation. Subjects took the least time to attain 100% capture for pacing with the Physio-Control device. No differences in performance times were seen with synchronized cardioversion among the devices. Human factors issues uncovered during this study included a preference for knobs over soft keys and a desire for clarity in control panel design. This study demonstrated no clearly superior defibrillator, as each of the models exhibited strengths in different areas. When asked their defibrillator preference, 67% of subjects chose the Philips.
Conclusions:
This comparison of user interfaces of defibrillators in simulated situations allows the assessment of usability that can provide manufacturers and educators with feedback about defibrillator implementation for these critical care devices.
More Related Videos
16:40A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
09:44Manual Muscle Testing: A Method of Measuring Extremity Muscle Strength Applied to Critically Ill Patients
Published on: April 12, 2011
Related Concept Videos
Cardiopulmonary Resuscitation III: AED Use
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Dysrhythmias VI: Management of Dysrhythmias
Cardiopulmonary Resuscitation I: Adult