Type IIIb Endoleak and Relining: A Mathematical Model of Distraction Forces
Charles Swaelens1, Robert J Poole2, Francesco Torella3
1Liverpool Vascular and Endovascular Service, Royal Liverpool University Hospital, Liverpool, UK swaelenscharles@gmail.com.
Purpose:
To examine the changes in distraction force following relining of a conventional abdominal aortic stent-graft with a type IIIb endoleak using the Nellix endovascular sealing device compared to a unilateral stent-graft.
Methods:
Relining is often used to repair type IIIb endoleaks, but the consequences to graft stability are unknown. A mathematical model was constructed based on pressure and volume flow through the stent-grafts, incorporating recognized distraction force equations. Steady flow was presumed at peak systolic pressures to calculate the maximum distraction force, with gravity ignored. Distraction forces for 28- to 36-mm-diameter stent-graft bodies with 16-mm limbs were calculated and compared to forces following relining with single and double Nellix devices or the Renu unilateral device.
Results:
Distraction forces for the 28-, 32-, and 36-mm stent-grafts prior to relining were 5.99, 10.21, and 14.99 N, respectively. Similar forces were reported after relining with bilateral Nellix devices (5.86, 10.08, and 14.86 N, respectively). However, use of a unilateral Nellix increased the distraction forces to 9.92, 14.14, and 18.92 N, respectively. These were comparable to the increase observed after relining with a Renu unilateral stent-graft (9.87, 14.09, and 18.86 N, respectively). The proportional increase in distraction force for a unilateral relining ranged from 26% to 66%, with the greatest increase noted in the smaller diameter main bodies.
Conclusion:
Relining a stent-graft with a type IIIb endoleak using bilateral Nellix devices does not increase the distraction force. However, a unilateral Nellix device or the Renu system could theoretically increase the distraction force by up to 66%, potentially risking migration and type Ia endoleak. In clinical practice, these results suggest that a relining with bilateral Nellix may have benefits over the Renu unilateral stent-graft.
More Related Videos
09:32Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
06:18Intravascular Ultrasound Image-Based Finite Element Modeling Approach for Quantifying In Vivo Mechanical Properties of Human Coronary Artery
Published on: December 6, 2024
Related Concept Videos
Application of the Linear Momentum Equation
The goal is to determine the force components in the x and y directions to hold the pipe in place. Since...
Design Example: Creating a Hydraulic Model of a Dam Spillway
Typical Model Studies
Design Example: Forces in Sluice Gate
Key variables in...
Relation between Poisson's ratio, Modulus of Elasticity and Modulus of Rigidity
Major Losses in Pipes
Fluid flow can be classified as laminar or turbulent, primarily based on the Reynolds number. This dimensionless number reflects the relative influence of inertial to viscous...