A Method for Systematic Electrochemical and Electrophysiological Evaluation of Neural Recording Electrodes
Published on: March 3, 2014
Verification of multichannel electrode array integrity by use of cross-channel correlations
Nicholas V Swindale1, Martin A Spacek1
1Department of Ophthalmology and Visual Sciences, University of British Columbia, Eye Care Centre, 2550 Willow St., Vancouver, BC, Canada V5Z 3N9.
Background:
The use of multichannel electrode arrays (MEAs) presents a number of practical challenges to experimenters including correctly labelling different recording channel locations and identifying sites that may be non-functional or short-circuited. These challenges are likely to increase as the number of sites used in recording increases.
New Method:
This paper presents a simple method for assessing MEA integrity based on the observation that physiologically induced signal correlations between nearby channels fall off with distance. Channels that violate this relationship are flagged as being potentially problematic.
Results:
The method is able to present to the user a list of potentially faulty channels for further inspection. Underlying problems include non-functional, shorted and mislocalised channels and channels carrying spurious noisy signals unrelated to those on other channels.
Comparison With Existing Methods:
Computational methods which automatically screen MEAs for faulty electrode channels do not appear to exist in the literature. Currently a user would have to examine single channels, or channel pairs, individually, which would be very time-consuming.
Conclusions:
Shorted or mislocalised channels may be more prevalent in MEA recordings than users suspect. The paper presents a simple screening method for identifying such channels prior to carrying out spike-sorting.
Related Concept Videos
Electrodes: Overview
There are two main types of electrodes in electrochemical cells. The first type, known as the working or indicator electrode, has a potential that is sensitive to the analyte's concentration and reacts to changes in...
Correlation between ECG and Cardiac Cycle
A cardiac action potential originates in the SA node and spreads throughout the atria and the AV node in approximately 0.03 seconds. This results in the P wave in an ECG and triggers atrial contraction. The action potential is then briefly slowed at the AV node, allowing the atria to contract and fill the ventricles with blood before...